Abstract

Remote unattended sensor networks are increasingly sought after to monitor the drinking water distribution grid, industrial wastewater effluents, and even rivers and lakes. One of the biggest challenges for application of such sensors is the issue of in-field device calibration. With this challenge in mind, we report here the use of anodic stripping coulometry (ASC) as the basis of a calibration-free micro-fabricated electrochemical sensor (CF-MES) for heavy metal determinations. The sensor platform consisted of a photo-lithographically patterned gold working electrode on SiO2 substrate, which was housed within a custom stopped-flow thin-layer cell, with a total volume of 2–4μL. The behavior of this platform was characterized by fluorescent particle microscopy and electrochemical studies utilizing Fe(CN)63−/4− as a model analyte. The average charge obtained for oxidation of 500μM ferrocyanide after 60s over a 10 month period was 176μC, corresponding to a volume of 3.65μL (RSD=2.4%). The response of the platform to copper concentrations ranging from 50 to 7500ppb was evaluated, and the ASC results showed a linear dependence of charge on copper concentrations with excellent reproducibility (RSD≤2.5%) and accuracy for most concentrations (≤5–10% error). The platform was also used to determine copper and mercury mixtures, where the total metallic content was measurable with excellent reproducibility (RSD≤4%) and accuracy (≤6% error).