Abstract
A silica monolithic capillary column was linked to an open capillary of the same internal diameter via a Teflon sleeve to form a duplex column to investigate the combination of chromatography and electrophoresis in the mode of electrically assisted capillary liquid chromatography (eCLC). Using a commercial CE instrument with an 8.5 cm long, 100 μm i.d. reversed phase silica monolithic section and a window 1.5 cm beyond the end of this in a 21.5 cm open section, a minimum plate height of 9 μm was obtained in capillary liquid chromatography (CLC) mode at a low driving pressure of 50 psi. In eCLC mode, high speed and high resolution separations of acidic and basic compounds were achieved with selectivity tuning based on the flexible combination of pressure (0–100 psi) and voltage. Taking advantage of the excellent permeability of silica monolithic columns, use of a step flow gradient enabled elution of compounds with different charge state.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.