Abstract

When integrating silicon nanowire biosensors with a microfluidic sample delivery system, additional challenges are introduced. Noise and erroneous signal generation induced by sample fluidic handling such as flow rate fluctuations during sample switching reduce the quality and reliability of the measurement system. In this paper, we propose an effective electrical shielding method to improve the stability and reliability of the setup by placing double electrodes instead of a single electrode that is traditionally used for nanowire sensors. Experimental results show that with proper shielding electrical measurements are not influenced by flow speed variations or during sample switching.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.