Abstract

In some applications, accurate knowledge of the dilution factor of a diluter can be crucial. Although the sample gas composition is one of the factors that influences the dilution factor of an ejector diluter, the available calibration data are usually obtained using air as sample gas. That can be a problem when the diluter is used in exhaust gas conditions, where the gas composition can differ significantly from air. Available ejector diluter models have been found unsuitable. The goal of this study was to set a method to determine the shift of dilution factor caused by sample gas composition, so the nominal or calibration value can be adjusted. Using CO2, the method was experimentally verified with good results. In addition, the present paper also investigated the possible influence of the diluter on the particle size distribution curve of the diluted sample.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.