1. These studies show that the readout error with a vibrating pointer is dependent on the operator, the position, and the state of the pointer. 2. Pointer vibrations of scale up to 0.2 division do not introduce an appreciable addition readout error. 3. Pointer vibrations with a scale of 0.3 division not only lead to an additional readout error of ±0.07 scale division but also increase the time required by 30%. 4. This method of examining readout error for a vibrating pointer can be used in examining various instruments used under conditions of vibration.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.