Abstract

Mechanical circulatory support is a potential treatment for failing Fontan patients. In this study, we performed a theoretical analysis using a computational model to clarify the effects of systemic ventricular assist device (VAD) in failing Fontan patients. Cardiac chambers and vascular systems were described using the time-varying elastance model and modified Windkessel model, respectively. A VAD was simulated as a nonlinear function. In systolic and diastolic ventricular dysfunction and atrioventricular valve regurgitation models, systemic VAD increased the cardiac index and decreased the central venous pressure (CVP). However, in the high pulmonary vascular resistance model, CVP became extremely high above 15 mmHg to maintain the cardiac index when the pulmonary vascular resistance index (PVRI) was above 5 Wood units m2. In Fontan patients with ventricular dysfunction or atrioventricular valve regurgitation, systemic VAD efficiently improves the hemodynamics. In Fontan patients with PVRI of > 5 Wood units m2, systemic VAD seems ineffective.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.