Abstract
Signal space separation (SSS) is a novel method for processing multichannel biomagnetic data. It is useful for a variety of applications including interference suppression, movement compensation and conversion of measurements between sensor arrays. The performance of SSS has been examined mainly on a 306-channel whole-head magnetoencephalography system. To facilitate the adaptation of the method to other biomagnetometer systems, the effect of various properties of the sensor array on its performance needs to be studied. To this end, we examined the effects of gradiometer imbalance, sensor calibration errors and erroneous sensor geometry information on SSS using simulations. The results indicate that depending on the application, gradiometer balance on the level of 0.1% to 0.5% may be needed for satisfactory SSS performance. For wire-wound gradiometers, this requires very careful attention in manufacturing. Errors in calibration coefficients and geometry information were found to have less significance.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.