Abstract

This study investigates the loss of compression when 3 commonly used headless compression screws are backed out (reversed), and assesses the ability to re-establish compression with screws of greater diameter. Two investigators tested 3 screw designs (Acutrak 2, Synthes HCS, Medartis SpeedTip CCS) in 2 diameters and lengths. Each design had 10 test cycles in a polyurethane foam bone model with compression recorded using a washer load cell. A 28-mm screw of the narrower diameter was inserted until 2 mm recessed and then reversed 30°, 60°, 90°, 180°, 270°, 360°, and 720°. After this the screw was removed completely and a 24-mm screw of greater diameter inserted until recessed 2 mm with the compressive force again recorded. All screws showed an immediate, statistically significant loss of compression at 30° of reversing. The Acutrak 2 Micro screw demonstrated not only the greatest mean compressive force, but also the fastest compressive loss. Insertion of the shorter screw of greater diameter was associated with re-establishment of compression to levels comparable with the original screw. This study reaffirms the importance of establishing the correct screw length before insertion due to the immediate loss of compression with reversal of these devices. If a headless compression screw penetrates the far joint surface, the screw should be completely removed and replaced with a shorter screw of greater diameter.