Abstract
Twenty-four insensible water loss (IWL) studies, measured as insensible weight loss (IL), were performed on 24 normal, term, newborn infants during the first two days of life. The IWL of infants placed under a radiant warmer (infrared radiant energy source) was significantly greater than the IWL of infants placed in a standard convection-type incubator (1.08±0.12 vs 0.53±0.05 gm/kg/hr; M±SE;P The higher IWL of infants in the radiant warmer is probably on the basis of a higher skin-ambient temperature gradient or a higher skin-ambient vapor pressure gradient, or both. These data illustrate the importance of providing proportionally larger fluid intake for maintenance in infants under the radiant warmer environment for thermal control.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.