Abstract

BackgroundThe validation of sample stability through pneumatic tube system (PTS) is essential. The objective of this study was to evaluate the effects of PTS transportation on laboratory results. MethodsPaired EDTA and SST blood samples were collected from 56 randomly selected patients. Laboratory parameters were compared between PTS group and hand-delivered group. ResultsNo statistical differences were observed for complete blood counts, white blood cell differential parameters, erythrocyte sedimentation rate and most chemistry parameters between PTS and hand-delivered transport procedures. Mean platelet volume results obtained from samples transported through PTS were lower than that obtained from samples transported through hand-delivered method (P = 0.001). The results of aspartate aminotransferase (P = 0.000), lactate dehydrogenase (P = 0.000), and hemolysis index (P = 0.000) from PTS group were higher than that from hand-delivered group. ConclusionsAll laboratories should validate the stability of the results from samples according to transportation method.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.