Abstract
The objective of the study was to find the incidence of hemolysis in samples transported through a pneumatic tube system (PTS) at different speeds. This prospective observational study was done in three phases: "short distance and high speed (115 m at 3 m/s)", "long distance and high speed (225 m at 3 m/s)" and "short distance and slow speed (115 at 2 m/s)". Fifty-two, 215 and 45 serum tube pairs, respectively, were evaluated in these three phases. A set of tubes was sent by PTS while the other was hand-carried. Samples were analyzed for supernatant hemoglobin (Hb), potassium (K+) and lactate dehydrogenase (LD). Mean transit time of samples through the PTS was much shorter as compared to human courier in all three phases. LD was elevated in PTS arm in the "short distance and high speed" phase and in the "long distance and high speed" phase, all three indices of hemolysis - Hb, K+ and LD - showed elevation in the PTS arm. However, at "short distance and slow speed" phase, there was no hemolysis in the PTS arm. Hospitals should validate their PTS before use and, by altering speed of sample transportation, hemolysis may be obliterated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.