Abstract

A molecular pump is always used as the main pump of the vacuum leak detection system because of its strong ability for pumping helium. But the pumping speed of a molecular pump is not enough for large vacuum chamber. A cryopump is a clean, oil-free and high pumping-speed vacuum pump. If a cryopump is used as the main pump of the vacuum leak detection system, the number of the pumps will decrease. This paper describes the experiments on vacuum leak detection system with cryopump as a main pump, and analyses the phenomena in the experiments. The results show that the cryopump decreased the vacuum pressure of the system significantly, but it had nearly no influence to the system sensitivity. The response time and the clean-up time of the leak detection system with only molecular pump were both 3 min. But the response time of the leak detection system with both molecular pump and cryopump was 10 min–20 min, and the clean-up time of the system was longer than an hour.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.