Abstract

BackgroundUnstable carotid plaques on subjective, visual, assessment using B-mode ultrasound scanning appear as echolucent and heterogeneous. Although previous studies on computer assisted plaque characterisation have standardised B-mode images for brightness, improving the objective assessment of echolucency, little progress has been made towards standardisation of texture analysis methods, which assess plaque heterogeneity. The aim of the present study was to investigate the influence of image zooming during ultrasound scanning on textural features and to test whether or not resolution standardisation decreases the variability introduced.MethodsEighteen still B-mode images of carotid plaques were zoomed during carotid scanning (zoom factor 1.3) and both images were transferred to a PC and normalised. Using bilinear and bicubic interpolation, the original images were interpolated in a process of simulating off-line zoom using the same interpolation factor. With the aid of the colour-coded image, carotid plaques of the original, zoomed and two resampled images for each case were outlined and histogram, first order and second order statistics were subsequently calculated.ResultsMost second order statistics (21/25, 84%) were significantly (p < 0.05) sensitive to image zooming during scanning, in contrast to histogram and first order statistics (4/25, 16%, p < 0.001, Fisher's exact test). Median (interquartile range) change of those features sensitive to zooming was 18.14% (4.94–28.43). Image interpolation restored these changes, the bicubic interpolation being superior compared to bilinear interpolation (p = 0.036).ConclusionTexture analysis of ultrasonic plaques should be performed under standardised resolution settings; otherwise a resolution normalisation algorithm should be applied.

Highlights

  • Unstable carotid plaques on subjective, visual, assessment using B-mode ultrasound scanning appear as echolucent and heterogeneous

  • The aim of the present study was to investigate the influence of image zooming during ultrasound scanning on the value of histogram analysis and textural features and to test whether or not resolution standardisation by applying image resampling decreases the variability introduced by the different image resolution

  • Eighteen images of carotid plaques producing stenosis greater than 50% were included in this study

Read more

Summary

Introduction

Unstable carotid plaques on subjective, visual, assessment using B-mode ultrasound scanning appear as echolucent and heterogeneous. Previous studies on computer assisted plaque characterisation have standardised B-mode images for brightness, improving the objective assessment of echolucency, little progress has been made towards standardisation of texture analysis methods, which assess plaque heterogeneity. The aim of the present study was to investigate the influence of image zooming during ultrasound scanning on textural features and to test whether or not resolution standardisation decreases the variability introduced. The aim of the present study was to investigate the influence of image zooming during ultrasound scanning on the value of histogram analysis and textural features and to test whether or not resolution standardisation by applying image resampling decreases the variability introduced by the different image resolution

Objectives
Methods
Results
Discussion
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.