Abstract

ObjectivesTo investigate in a numerical study the effect of the geometry and the extension of orthodontic aligner edges and the aligner thickness on force transmission to upper right central incisor tooth (Tooth 11). MethodsA three-dimensional (3D) digital model, obtained from a 3D data set of a complete dentulous maxilla, was imported into 3-matic software. Aligners with four different trimming line designs (scalloped, straight, scalloped extended, straight extended) were designed, each with four different thicknesses (0.3, 0.4, 0.5, and 0.6 mm). The models were exported to a finite element (FE) software (Marc/Mentat). A facial 0.2 mm bodily malposition of tooth 11 was simulated. ResultsThe maximum resultant force was in the range of (7.5 - 55.2) N. The straight trimming designs had higher resultant force than the scalloped designs. The resultant force increases with increasing the edge extension of the aligner. The normal contact forces were unevenly distributed over the entire surface and were concentrated in six areas: Incisal, Mesio-Incisal, Disto-Incisal, Middle, Mesio-Cervical, and Disto-Cervical. The resultant force increases super linearly with increasing thickness. ConclusionsThe design of the trimming line, the edge extension, and the thickness of the aligner affect significantly the magnitude of the resultant force and the distribution of normal contact force. The straight extended trimming design exhibited better force distribution that may favor a bodily tooth movement. Clinical relevanceA straight extended trimming design of an orthodontic aligner may improve the clinical outcomes. In addition, the manufacturing procedures of the straight design are much simpler compared to the scalloped design.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call