Abstract
In “comprehensive” two-dimensional liquid chromatography, the column effluent from the first separation system (the first dimension) is sequentially sampled by the second dimension separation system. The total analysis time is largely determined by the speed of the second dimension separation system; the most retained component must elute before the least retained component of the next second dimension separation. Optimization of multidimensional separation systems requires that one understand the relationship between system resolution and the number of second dimension samples across a first dimension peak. In this paper, we study the theoretical and experimental aspects of this sampling process. To obtain high two-dimensional resolution, each peak in the first dimension should be sampled at least three times into the second dimension when the sampling is in-phase. If the sampling is maximally out of phase, there should be at least four samples per peak for high-fidelity separation. The sensitivity of th...