Abstract

The aim of this study was to evaluate how preset torque settings influence the torque, vertical force, and root canal-centering ability of ProGlider and ProTaper NEXT nickel-titanium rotary instruments in canals with different curvature locations. Based on micro-computed tomography, mesial roots of human mandibular molars (25°-40° curvature) were allocated to the apical curvature (apical 1-5 mm) or the middle curvature (apical 5-9 mm) groups, and mandibular incisors (curvature <5°) to the straight canal group. Each group was subjected to automated instrumentation and torque/force measurement with the preset torque of 1, 2.5, or 5 N•cm. Canal-centering ratios were determined with micro-computed tomography. Instrument fracture occurred only in the 2.5 and 5 N•cm groups in curved canals. The preset torque setting and curvature location did not influence canal shaping ability.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call