Abstract
Serial (baseline and 9-month follow-up) intravascular ultrasound analysis was performed at 5-mm reference segments immediately proximal and distal to the sirolimus-eluting stent (SES) in 33 lesions. Proximal and distal reference segments were divided into 1-mm subsegments. Between postintervention and follow-up intravascular ultrasound studies, there were significant decreases in the lumen and increases in plaque & media areas in the subsegment closest to the distal edge, with no change in external elastic membrane area. There was no significant change in external elastic membrane, lumen, and plaque & media areas within the other subsegments. At the nearest 1-mm subsegment from the proximal and distal edges, baseline plaque & media area was associated with subsequent vessel remodeling. In conclusion, a large amount of plaque at the SES edge may be a risk of negative remodeling at follow-up (stent edge restenosis). It supports the importance of "normal-to-normal" SES deployment.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.