Abstract

We aimed to evaluate the quality of chest computed tomography (CT) images obtained with low-dose CT using three iterative reconstruction (IR) algorithms. Two 64-detector spiral CT scanners (HDCT and iCT) were used to scan a chest phantom containing 6 ground-glass nodules (GGNs) at 11 radiation dose levels. CT images were reconstructed by filtered back projection or three IR algorithms. Reconstructed images were analyzed for CT values, average noise, contrast-to-noise ratio (CNR) values, subjective image noise, and diagnostic acceptability of the GGNs. Repeated-measures analysis of variance was used for statistical analyses. Average noise decreased and CNR increased with increasing radiation dose when the same reconstruction algorithm was applied. Average image noise was significantly lower when reconstructed with MBIR than with iDOSE4 at the same low radiation doses. The two radiologists showed good interobserver consistency in image quality with kappa 0.83. A significant relationship was found between image noise and diagnostic acceptability of the GGNs. Three IR algorithms are able to reduce the image noise and improve the image quality of low-dose CT. In the same radiation dose, the low-dose CT image quality reconstructed with MBIR algorithms is better than that of other IR algorithms.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.