Abstract

We investigated how the distribution of magnitude and phase of incident electric field affects RFinduced heating near a hip implant. The results showed that varying the incident electric field, for example due to different phantom shape or different landmark position, for two-or three-dimensional implants can result in up to 50% variation of estimated RF-induced temperature rise. To avoid systematic errors in predicting the RF-induced heating, varied distributions of the incident electric field should be applied.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.