Abstract

The purpose of the study is to evaluate the effect of energy level on the modulation transfer functions (MTF) and noise power spectra (NPS) of virtual monochromatic images (VMIs) obtained using four types of computed-tomographic (CT) scanners: Revolution, SOMATOM, IQon, and Aquilion. VMIs were obtained at 70, 60, and 50 kiloelectron volts (keV), and also at the lowest keV available in each scanner. We evaluated the MTF and NPS in the VMIs obtained at each keV. No significant effect of the energy level on the MTF was observed in IQon, whereas the spatial resolution decreased as the energy level decreased in the other types of scanners. The NPS curves tended to increase as the energy levels decreased with three types of scanners other than Aquilion. The spatial resolution and noise frequency characteristics of VMIs may be affected by the energy level, and the effects of energy level on these characteristics differ depending on the type of CT scanners.