Abstract
In x-ray imaging, high resolution is essential, particularly in sectors such as medical and industries where the need for nondestructive defect detection is required. Previous research has shown that altering beam design and the number of gates offset holes has an impact on focal spot size (FSS). However, the specific effects of beam size and offset size were not thoroughly assessed. In the present study, the influence of beam size and gate offset size was evaluated by utilizing a cold cathode electron beam. Various beam sizes were employed to achieve a small FSS, and subsequently, the smallest beam was utilized to examine the impact of gate offset size. In doing so, the smallest FSS of 0.25 mm vertical and 0.33 mm horizontal was attained without the utilization of any additional focusing lens. This illustrates that by adjusting the beam size and gate offset size, it is possible to attain a small FSS, facilitating the development of an economically viable x-ray imaging beam.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.