Abstract
We report on the structural effects of a carbon nanotube based cold cathode electron beam (C-beam) module on focal spot sizes (FSSs) for high resolution x-ray imaging applications without additional focusing components. A simple combination of vertically aligned carbon nanotubes and gate mesh structure achieved FSS of 252 and 410 μm for longitudinal and transverse directions, respectively. As design parameters of the modules, gate offsets, gate hole pitch, and emitter area effects of C-beam modules on FSS were optimized. We expect that our simple but sophisticated C-beam module structure will pave next generation x-ray devices not only for high quality x-ray imaging but cost-effective ways for mass production.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.