Abstract
The aim of this study was to evaluate the pullout strength of 3 different orthodontic mini implants. Twenty-seven mini implants (diameter: 2 mm, length: 7 mm) were implanted into artificial bone (Sawbones; Pacific Research Laboratories Inc.) at depths of 3, 4, and 5 mm. The insertion torque (IT), resonance frequency (RF), pullout strength (PS), and anchor length (AL) were measured. One-way analysis of variance with Tukey honest significant difference (HSD) postcomparison were used to detect intergroup differences. The null hypothesis was that IT, RF, and PS would significantly correlate in the same brand. In the implantation depths (ID) (5 and 4 mm), IT of Types C (16.67 and 14.33 N·cm) and Type B (14 and 13.33 N·cm) were significantly higher than Type A (10.33 and 9.33 N·cm). Type A had a largest AL and PS at the IDs (5 and 4 mm). In the IDs (3 mm), PS was no different. Type C had no correlation among the RF, IT, and PS. Therefore, null hypothesis was rejected. AL exerted crucial effects on the PS of the mini implants.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.