Abstract

The effect of using ammonia as a carrier gas on the response of the flame ionization detector (FID) has been investigated. It was found that the FID response, calculated as the effective carbon number (ECN), increased for all the compounds studied when ammonia, rather than helium, was used. The change was 0–0. 9 carbon atom for hydrocarbons, one carbon atom for alcohols and diphenyl ether, and 0.4–1 carbon atom for phenols and ketones. The increase in ECN was larger for amines (0. 8–5 carbon atoms), but these numbers also reflected an improvement in chromatographic performance as a result of reduced adsorption on the column. The largest change in signal-to-noise ratio, a six-fold increase, was obtained for octyl-amine; ratios for hexyl methyl ketone, diisobutyl ketone, dihexyl-amine, dibutylamine, and N-methyloctylamine increased by a factor of 2–3 when ammonia was used as carrier gas. To determine the extent to which the effect on detector response was solely attributable to ammonia, a mixture of 5 % ammonia in nitrogen was used as detector make-up gas with helium as carrier gas. Under these conditions the noise in the FID increased but for most of the compounds studied the signal-to-noise ratio also increased.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.