Abstract

This study aimed to compare the radiofrequency (RF) shielding effects of titanium mesh of echo-planar imaging (EPI) versus fast spin-echo (FSE) diffusion-weighted imaging (DWI) to establish a suitable sequence for patients who undergo cranioplasty and for whom titanium mesh was used in brain magnetic resonance imaging (MRI). A 1.5-T MRI scanner with clinical setting sequences was used. A phantom for the examination constructed using a sucrose solution in a plastic container was used to compare the signal attenuation (SA) ratio, area of RF shielding effect (Area), normalized absolute average deviation (NAAD), and apparent diffusion coefficient (ADC) between EPI and FSE-DWI. EPI provided significantly better SA ratio, Area, and NAAD (P < 0.01). When the number of slices increased, the RF shielding became more negative. There was no significant difference in the ADC. Regardless of the k-trajectory, EPI-DWI had a lower RF shielding effect than FSE-DWI in patients undergoing cranioplasty.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.