Abstract

Confocal microscopes use photomultiplier tubes and hybrid detectors due to their large dynamic range, which typically exceeds the one of single-photon avalanche diodes (SPADs). The latter, due to their photon counting operation, are usually limited to an output count rate to 1/Tdead. In this paper, we present a thorough analysis, which can actually be applied to any photon counting detector, on how to extend the SPAD dynamic range by exploiting the nonlinear photon response at high count rates and for different recharge mechanisms. We applied passive, active event-driven and clock-driven (i.e. clocked, following quanta image sensor response) recharge directly to the SPADs. The photon response, photon count standard deviation, signal-to-noise ratio and dynamic range were measured and compared to models. Measurements were performed with a CMOS SPAD array targeted for image scanning microscopy, featuring best-in-class 11 V excess bias, 55% peak photon detection probability at 520 nm and >40% from 440 to 640 nm. The array features an extremely low median dark count rate below 0.05 cps/μm2 at 9 V of excess bias and 0°C. We show that active event-driven recharge provides ×75 dynamic range extension and offers novel ways for high dynamic range imaging. When compared to the clock-driven recharge and the quanta image sensor approach, the dynamic range is extended by a factor of ×12.7-26.4. Additionally, for the first time, we evaluate the influence of clock-driven recharge on the SPAD afterpulsing.

Highlights

  • Confocal microscopes use photomultiplier tubes (PMTs), hybrid detectors [1] or singlephoton avalanche diodes (SPADs)

  • We present a thorough analysis, which can be applied to any photon counting detector, on how to extend the SPAD dynamic range by exploiting the nonlinear photon response at high count rates and for different recharge mechanisms

  • Systems intended for image scanning microscopy (ISM) applications employ either a combination of discrete sensors, or linear arrays coupled through fibers and microlenses in order to form a 2D array at the microscopy output [7]

Read more

Summary

Introduction

Confocal microscopes use photomultiplier tubes (PMTs), hybrid detectors [1] or singlephoton avalanche diodes (SPADs). We examine the possibility and limitations of extending the SPAD dynamic range, for different recharge mechanisms, by utilizing the exponential distribution of photons in time, which results in a nonlinear photon response This methodology can be extended to any photon counting detector.

CMOS high-performance SPAD array
Afterpulsing and response linearity with clock-driven recharge
Findings
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.