Abstract
Pressure transducers are mainly used for dynamic measurements but calibrated by means of static calibration methods. The reason is that neither a standardized method for a traceable dynamic secondary calibration of pressure transducers nor a primary dynamic calibration method for reference pressure transducers is available. The only exceptions are the calibration methods for pressure field microphones which can also be regarded as dynamic air pressure transducers for the low pressure range. This paper compares results from measurements with two different pressure exciters that have their origin in acoustics and were used for calibration of two types of pressure transducers in the frequency range 31.5 Hz to about 1 kHz at pressure amplitudes up to some 100 Pa. In both cases, a traceably calibrated pressure microphone was used as reference sensor that should provide a traceability of the pressure transducer calibration. The paper discusses the pros and cons of both exciters and if this can be considered as a traceable calibration method.
Highlights
Pressure transducers are mainly used for dynamic measurements but calibrated by means of static calibration methods
The only exceptions are the calibration methods for pressure field microphones which can be regarded as dynamic air pressure transducers for the low pressure range
A traceably calibrated pressure microphone was used as reference sensor that should provide a traceability of the pressure transducer calibration
Summary
Pressure transducers are mainly used for dynamic measurements but calibrated by means of static calibration methods. The reason is that neither a standardized method for a traceable dynamic secondary calibration of pressure transducers nor a primary dynamic calibration method for reference pressure transducers is available. The only exceptions are the calibration methods for pressure field microphones which can be regarded as dynamic air pressure transducers for the low pressure range. The paper discusses the pros and cons of both exciters and if this can be considered as a traceable calibration method
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.