Abstract

Existing dynamic mechanical testing systems are not well-suited for Raman spectroscopic study of the response of bone tissue chemical structure to mechanical loading. This paper describes design and validation of a testing system for this application. The body of the apparatus is waterproof and is closed to allow passive maintenance of tissue hydration. Specialized specimen grips for tension/compression and for bending studies are described. The system is shown to provide reproducible data on test specimens of murine femora and radii. Stability against unstable grips and long term (12 h) stability against dehydration artifacts are demonstrated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.