Abstract

Whilst contact angle measurements obtained using the Wilhelmy balance technique are accurate and reproducible for planar surfaces, their use for characterizing particulate materials is highly dependent upon accurate knowledge of the wet perimeter. This communication suggests that the approach of Pepin et al. [Int. J. Pharm. 152 (1997) 1] for wet perimeter determination using non-polar liquids may lead to erroneous conclusions. Alternative approaches for wet perimeter determination are suggested.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.