Abstract

In optical spectroscopic systems where unwanted optical scattering cannot be eliminated, Fabry-Pérot etalons cause unpredictable changes in the spectral background. Frequent system calibration is then required to maintain the desired measurement accuracy, which presents a major limitation to the spectrometer. We introduce a computational approach to mitigate the adverse effects of optical fringing without hardware modifications. Motivated by experimental observations of complicated fringe behaviors, we simplify the problem by decomposing the fringe background into component etalons that can be addressed according to their individual characteristics. The effectiveness of the proposed method is demonstrated on a silicon photonic methane sensor, where accurate measurements of methane concentration are obtained from spectral data strongly affected by optical fringes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.