Abstract

Dynamic calibration is one of the important ways to obtain the dynamic performance parameters of pressure sensors, and shock tube calibration is a common method for dynamic calibration of pressure sensors. To study the dynamic calibration method of the shock tube of the blast pressure pencil probes, such as free-field pressure sensors and dynamic pressure sensors, a convenient and adjustable PVC-U (unplasticized polyvinyl chloride) shock tube is developed. Taking the dynamic pressure sensor as the research object, the feasibility of dynamic calibration of shock tube based on CFD (Computational Fluid Dynamics) techniques is discussed. Finally, the dynamic calibration experiment based on shock tube is carried out, and the time-domain and frequency-domain signals are analyzed. The results are in good agreement with the theoretical analysis. The results show that the shock tube can effectively excite the natural frequency of the dynamic pressure sensor, and the developed shock tube can indeed achieve the dynamic calibration and dynamic experiments of pencil probes.