Abstract

This paper provides new geometry definitions for the axi-symmetric stiffening (or sewing) ring, as part of a bi-leaflet prosthetic heart valve, and presents a dynamic behaviour analysis of the leaflet. An optimal stiffening ring geometry may be constructed by considering the point of flow separation on the stiffening ring (measured in the downstream coordinate), the effective orifice area of the stiffening ring and the associated dynamic behaviour of the leaflet. The dynamic model is accomplished by utilizing a second-order rotating system to simulate the opening and closing characteristics of the leaflet. The moments due to the aerodynamic loads are evaluated from an irrotational inviscid flow model, coupled with boundary layer theory, modelling the internal flow phenomena of the bi-leaflet heart valve implanted in the aortic root. It has been shown that this internal flow model provides the correct evaluation of lift and induced drag and the subsequent dynamic characteristics.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.