Abstract

A dual stage preconcentration system based on flow injection on-line ion-exchange and solvent-extraction has been developed for flame atomic absorption spectrometry. Lead is taken as a model trace element. A column packed with Amberlite IRC-718 cation exchanger is incorporated into the FI manifold. The analyte is retained on the column by time-based sample loading and eluted by 1 mol/L HNO3. The eluate is subsequently merged with potassium iodide and tetrabutylammonium bromide (TBABr), and isobutyl methyl ketone (IBMK). Lead is extracted on-line into IBMK as the ion-pair formed between the iodoplumbate anion and tetrabutylammonium cation. The organic phase is separated from the aqueous phase by a gravity phase separator. 50 μL of concentrate is introduced into the nebulizer-burner system of the spectrometer. An enhancement factor of 550 is achieved with a 30 mL sample consumption at a sampling frequency of 30/h. The precision (relative standard deviation) is 2.4% at 10 μg/L level and the detection limit is 0.3 μg/L (3 σ). The method was successfully applied to the determination of lead in water samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.