Abstract

It is evident that surface electromyography (sEMG)-based sensing approach for human-machine interfaces has some inherent limitations for applications requiring morphological changes information of deep-layered muscles, such as dexterous prosthetic hands. In this paper, the design, simulation, fabrication, and evaluation for a series of novel structured ultrasound transducers are conducted in order to develop a type of A-mode ultrasound transducers that overcome the drawbacks of the sEMG-based sensing. The transducers cover single-frequency and dual-frequency types. Their key parameters, the acoustic impedance and thickness of the matching layer, are simulated and verified by PZFlex. The parameters are designed as 0.3 times of the 1-3 composite piezoelectric's acoustic impedance and 0.25 times of the wavelength, respectively. The characterizations of the dual-frequency transducers significantly outperform single-frequency transducers. The experiments of recognizing dexterous hand gesture are designed to detect morphological changes information of deep-layered muscles. The classification accuracy improvements with linear discrimination analysis are 7.3% and 4.7%, and with support vector machine are 14.1% and 13.4% for the horizontal stacked and annulus array. This preliminary study concludes that the dual-frequency transducers have huge potential for applications that need contraction information of deep-layered muscles over the single-frequency transducers, letting alone sEMG-based sensors.