Abstract

A reagent strip for quantifying theophylline in serum or plasma has been developed for use with the Apoenzyme Reactivation Immunoassay System (ARIS) with the Ames Seralyzer reflectance photometer. The test takes 80 s and involves comparison with a two-point calibration line, which can validly be stored in the instrument for two weeks. Results for theophylline in clinical serum samples correlate well (r greater than 0.98) with results by liquid chromatography, fluoroimmunoassay, and enzyme immunoassay procedures. The within-run CV for four concentrations of controls ranged from 3.5 to 6%; the between-run CVs ranged from 3 to 5%. This assay for use in therapeutic drug monitoring is convenient, rapid, and simple, and thus is appropriate for use in emergency rooms, physician's offices, and small laboratories.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.