Abstract

A reagent strip for the quantitative analysis of theophylline in serum or plasma was evaluated. The strip is based on the apoenzyme reactivation immunoassay system (ARIS) technique and is intended for use with the Ames Seralyzer reflectance photometer. The method gave CVs at three theophylline levels ranging from 3.8 to 6.3% (within run) and from 2.8 to 6.9% (day to day). The regression lines obtained from the correlation studies were y = 0.959x + 0.51 (n = 105, r = 0.9906, Sy/x = 0.56) for the comparison ARIS (y) versus Syva enzyme multiplied immunoassay (x) methods, and y = 0.986x + 0.32 (n = 105, r = 0.9832, Sy/x = 0.62) for the comparison ARIS (y) versus Abbott TDx fluorescence polarization immunoassay (x) methods. The interference from triglycerides, hemoglobin, bilirubin, and ascorbic acid, and the cross-reactivity of 8-chlorotheophylline, caffeine, 1,3-dimethyluric acid, theobromine, and 1,7-dimethylxanthine, were also investigated and discussed. The method was found to be reliable, simple, and rapid. It provides a practicable solution for immediate determinations of theophylline.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.