Abstract

Nebulizer drug output rate increases during the nebulization. For unvented jet nebulizers, a physical and mathematical model based on the efficiency of the nebulization process is presented for this phenomenon. Formulas are derived for the cumulative drug output and the drug output rate of the nebulization process. The model is compared with the model proposed by Coates et al. [J. Aerosol. Med. 11 (1998) 101]. Both models are supported by experimental literature data. Both models predict the experimental values well but the proposed model allows more easy prediction of the influence of small changes in the nebulization conditions and the calculation of the cumulative drug output for a related process. From literature data it is shown that the efficiency of an unvented jet nebulization process of diluted aqueous solutions is relatively insensitive to small changes in the concentration as well as to small changes in aspiration flow but is sensitive to the humidity of the compressor gas only.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.