Abstract

Oscillatory and pulsatile fluid flows for use in microfluidic applications were generated using a deformable chamber driven by a low cost linear voice coil actuator. Compliance in the fluidic system originating in the deformable chamber and the fluidic tubing produced a strong frequency dependence in the relationship between the system's input and the output flow rate. The effects of this frequency dependence were overcome by precise system calibration, enabling on-demand generation of sinusoidal oscillations in the fluid flow rate with a controlled amplitude in the range from 0.1 to over 1 ml/min across a frequency range from 0.1 Hz to 10 Hz. The calibration data further enabled the optimization of a multistage exponential smoothing model of the system that allowed the generation of arbitrary complex waveforms. This was demonstrated by combining flow modulation with a constant background flow generated by a syringe pump to mimic the pulsatile flow found in the human vascular system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.