Abstract

The effects of drifting instrument sensitivity in glow discharge optical emission spectrometry (GD-OES) on instrument calibration were investigated. It was found that for some analytical applications, the rate of drift may be fast relative to the time needed to calibrate the instrument. The results indicate that if this is the case, the validity of calibration may be compromised, introducing analytical bias. A new drift correction method recently published elsewhere was shown to eliminate any drift-induced calibration difficulties and the associated potential for analytical error. This is important, because drift correction methods that have been traditionally used in OES for decades are not applicable to this particular type of drift. Therefore, the new drift correction method should enhance the utility of GD-OES for important applications, such as quality control and quality assurance in industry.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.