Abstract
We developed new forward-viewing matrix transducers consisting of double ring arrays of 118 total PZT elements integrated into catheters used to deploy medical interventional devices. Our goal is 3D ultrasound guidance of medical device implantation to reduce x-ray fluoroscopy exposure. The double ring arrays were fabricated on inner and outer custom polyimide flexible circuits with inter-element spacing of 0.20 mm and then wrapped around an 11 French (Fr) catheter to produce a 15 Fr catheter (outer diameter [O.D.]). We used a braided cabling technology to connect the elements to the Volumetrics Medical Imaging (VMI) real-time 3D ultrasound scanner. Transducer performance yielded an average -6 dB fractional bandwidth of 49% ± 11% centered at 4.4 MHz for 118 elements. Real-time 3D cardiac scans of the in vivo pig model yielded good image quality including en face views of the tricuspid valve and real-time 3D guidance of an endo-myocardial biopsy catheter introduced into the left ventricle.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.