Abstract

The dose uniformity of anterior chest treatments resulting from a missing tissue compensating filter has been studied. Patient dosimetry with 10 MV X-rays was simulated using film, an ionization chamber, and thermoluminescent dosimetry (TLD). The results show that the compensating filter greatly improved the dose uniformity of a homogeneous phantom. However, the improvement for an inhomogeneous phantom and for patients was found to be less impressive. When using the filter, the increase in surface dose was found to be clinically insignificant.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.