Abstract

Because of reproducibility and repeatability problems with reference measurements made with an ionisation chamber in a self-protected GSRD1 irradiator, a comparison was made with alanine dosimetry for a whole-body mouse irradiation setup in a sterile box. The twisting of the cables in the cable duct and in the irradiator cell and the irradiation of the ionisation chamber connector are likely to have caused the problems encountered. These problems are not observed on other types of irradiators with more suitable cable passages. A difference up to 8.4% was observed between the alanine dosimetry and ionisation chamber. The influence of the number of animals in the sterile box on the whole-body dose of the animals was also evaluated with alanine and found to be <2%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.