Abstract

A commercially available variable electron beam applicator designed for the Phillips SL-75/20 linear accelerator was tested for possible clinical use. The basic design of this collimator consists of four movable L-shaped stainless steel blades forming the corners. In addition, four smaller fixed blades of the same material and thickness fill in the sides. Film placed in polystyrene phantoms were exposed in all experiments. Field size, field flatness, beam homogeneity and penumbra measurements, all indicated that the applicator is acceptable for clinical use. However, for use with large field sizes and/or high energy electron beams some minor modifications are suggested.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.