Abstract

The dosimetry of high-energy photon beams in the treatment of superficial lesions occurring in extremities was examined. Large parallel-opposed fields with different photon beam energies were used. The extremity was immersed in water contained in a commercially available plastic wastebasket. The water bolus serves to even out the surface irregularities of the extremities and to remove the skin sparing effect. A polystyrene block was placed at the floor of the wastebasket to ensure that the extremity was encompassed in the radiation fields. The photon beam energies considered were 4 MV, 6 MV, 10 MV, and 24 MV. The results show that the dose distributions are more homogeneous with higher photon beam energies. The isodose lines are more constricted at mid-plane for low energy photon beams. Higher energy photon beams, 10 MV and up would be preferable for the treatment of superficial lesions of the extremities immersed in water bolus contained in a typical wastebasket size.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.