Abstract

Physical wedges (PWs) are widely used in radiotherapy to obtain tilted isodose curves, but they alter beam quality. Dynamic wedges (DWs) using moving collimator overcome this problem, but measuring their beam data is not simple. The main aim of this study is to obtain all dosimetric parameters of DWs produced by Varian 2100CD with Monte Carlo simulation and compare them to those from PWs. To simulate 6 MV photon beams equipped with PW and DW, BEAMnrc code was used. All dosimetric data were obtained with EDR2 films and two-dimensional diode array detector. Good agreement between simulated and measured dosimetric data for PW and DW fields was obtained. Our results showed that percentage depth dose and beam profiles at nonwedged direction for DWs are the same as open fields and can be used to each other. From Monte Carlo simulations, it can be concluded that DWs in spite of PW do not have effect on beam quality and are good options for treatment planning system which cannot consider hardening effect produced by PWs. Furthermore, BEAMnrc is a powerful code to acquire all date required by DWs.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.