Abstract

One of the major sources of the coefficient of variation in quantitative digital image analysis systems is the distribution error. It is caused by uneven optical density distribution within the measuring area. In digital image analysis systems, the measuring area is an individual pixel. The distribution error would be eliminated if we could measure the average optical density within the measuring area. However, most cameras used with light microscopes measure the average transmittance of a pixel. The image analysis system then converts the measured transmittance into an optical density. If the optical density distribution within a pixel is not uniform, the average optical density converted from the measured transmittance is different from the true optical density of the pixel. The authors call this conversional distribution error (CDE). They have analyzed and characterized this error, and developed an algorithm to minimize CDE by estimating the optical density distribution within a pixel. >

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.