Abstract

Abstract This work describes the preparation of mesoporous carbon ceramics SiO 2 /50 wt.% C ( S BET = 170 m 2 g −1 ), where C is graphite, by the sol–gel method. This material was used as a matrix to support cobalt phthalocyanine (CoPc), prepared in situ on their surface, to assure homogeneous dispersion of the electrocatalyst complex in the pores of the matrix. Pressed disk electrode made with SiO 2 /C/CoPc was used to analyze dissolved oxygen in water by an electrochemical technique in neutral media and dissolved O 2 was reduced at −0.23 V. The linear response range, sensitivity and detection limit obtained were 0.5–6.6 mg L −1 , 2.16 μA L mg −1 and 0.01 mg L −1 , respectively, by chronoamperometry. A mechanism involving two electrons in O 2 reduction was determined by cyclic voltammetry technique. The repeatability of the proposed sensor, evaluated in terms of relative standard deviation was 1.6% for 10 measurements of a solution of 6.5 mg L −1 oxygen. The sensor was applied to determine oxygen in pond and tap water samples showing to be a promising tool for this purpose.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.