Abstract

AbstractThe present manuscript describes the development of a sensitive and selective disposable solid‐state potentiometric sensor based on doped polypyrrole films (PPy) for dipyrone determination. The dipyrone was incorporated in the PPy matrix during electrochemical polymerization realized under galvanostatic conditions at a graphite pencil electrode. The sensor response for dipyrone was linear in the concentration range from 1.0×10−4 to 4.0×10−2 mol L−1 with a slope of 29 mV dec−1 without interference of inorganic and/or organic species. The method was successfully applied in the determination of dipyrone in spiked human urine and pharmaceutical commercial samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.