Abstract

In order to improve the service life of the artificial acetabular cup in a total hip replacement, it is important to determine the best material and design, and to assess the mechanical behavior around the cup. In this study, electronic speckle interferometry (ESPI) and the two-dimensional finite element method (FEM) are employed to investigate the mechanical behavior. The influence of the cancellous bone and cup thickness on mechanical behavior around the cup was investigated. Good agreement of the cup model was found between the ESPI measurements and FEM predictions. The following results were obtained. (1) Cancellous bone with a porous structure can be measured by the ESPI method. (2) There are discontinuities of the displacement distribution in the transverse direction in each boundary region of the cup, bone cement and cancellous bone. (3) The maximum shear stress exists in the boundary region of the cup and bone cement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.