Abstract
The effective disinfection of hospital surfaces is recognised as an important factor in preventing hospital-acquired infections. The purpose of this study was to quantify the disinfection rate of a novel gas plasma system on clinically relevant biofilms. Clinical isolates of Staphylococcus epidermidis and meticillin-resistant Staphylococcus aureus (MRSA) were grown as biofilms on glass surfaces and tested in a disinfection container remote from the plasma source. The strains used in this study were known to produce substantial quantities of biofilm and average log₁₀ counts were 9.0 and 9.1 cfu/cm(2) for S. epidermidis and MRSA respectively. Counts were reduced by between 4 and 4.5 log₁₀ after 1h of exposure for MRSA and S. epidermidis respectively. More prolonged treatment in the case of MRSA biofilms resulted in a 5.5 log₁₀ reduction after 90 min. Biofilm samples were also placed in medical device packaging bags and similar rates of disinfection were observed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.